Welcome to Optical Comparators blog.

Since 1986 Dorsey Metrology have been manufacturing high quality optical comparators.

You can contact us about optical comparators sales, service and general informaion by calling directly to +1-845-454-3111.

Archive for 'manuals'

Optical comparator calibration

Squareness of X axis to Y axis

Squareness of X axis to Y axis is checked by placing a square on the center of the optical comparator table banked up against one of the stage vees. Face it to the right or left side. Place a magnetic base indicator on the face of the comparator near the chart ring. Lower the Y-axis to the bottom and place the tip of the indicator on the side of the square. Raise the Y-axis and check the squareness of the X axis in relation to the Y axis travel. It should be .0001” or less per inch (≤ .0006” for 6” travel). If adjustment is needed the knee bracket must be loosened and twisted to correct.

Squareness of Y axis to focus plane

Squareness of Y axis to focus plane is checked in the same manner as squareness of X axis to Y axis only with the face of the square facing forward or back. This should be .0005” per inch or less (≤ .003” for 6” travel). If this is not correct you would need to shim under the lower focus bearings to bring into spec.

X axis stage linear traveling accuracy

For this specification we use the formula ± (150+L/.02)µin. For a 10” travel optical comparator stage we would divide 10” by .02 to get 500 then add to it 150 for a ± total of 650 micro inches over 10” travel (±.00065”). The optical comparator stage is inspected using the 6” scale on the Acu-Rite master in the center of the stage. The deviation from the expected reading is recorded on the inspection sheet in micro inches (.000001”), a minus sign before the value indicates the measurement was under size. If corrections are needed a new comp. value can be entered into the optical comparator digital readout for the axis you are working on. Errors that are linear in nature are easily compensated for. If your readings are very non-linear it could indicate damage or looseness. The remarks section of the inspection sheet shows any error compensation set in the digital readout for that axis.

Y axis stage linear traveling accuracy is checked the same as X only with the master standing vertical on the stage.

ERP calibration counts

This value indicates the number of pulses recorded by the rotary encoder during calibration. The idea is to have a point of reference to use in the future if any errors or changes are needed to make the digital angle display more accurate. To re-calibrate the rotary encoder follow the instruction in the optical comparator manual and record the new value on the inspection sheet after a satisfactory value is achieved on the display.

ERP Repeatability

A check in this box indicates that the angle display is within ± 30 minutes of angle over a 360° rotation (±0.5°).

Replacing the stage on optical comparator

First free the two scale wires that run to the digital readout and coil them up and secure with a rubber band or wire twist tie.

Move the stage to the extreme left and right to enable removal of the four stage mounting screws.

Open the two lead screw quick releases and lift the stage off of the comparator holding it horizontal. CAUTION: If it is tilted it will slide quickly to one side or front to back and might fall or get damaged.

Take note: If there are any shims or washers under the stage at the mounting location. If shims or washers are present they will have to be replaced the same way they are located now.

Place the new stage on the comparator and then remove the shipping locks. CAUTION: If you tilt the stage during installation it will slide to the side and will damage the teeth in the quick release that engages the lead screw or will damage the stage.

Secure the stage to the optical comparator with the four mounting screws. Leave three of the screws loose and one just snug so that the stage can pivot around it. Turn the screen so that the vernier scale is at 0 or the screen pointer indicates that the horizontal line is set to its horizontal position. Place a pointer on the optical comparator stage so that it appears on the screen along the horizontal line. Now using the lowest magnification lens move the stage left and right using the quick release. If the pointer does not follow the screen horizontal line twist the stage so that the left and right stage travel matches the horizontal line. Then tighten all stage mounting screws and re-attach the scale cables.

Use the locking hardware removed from the new optical comparator stage to lock the old stage for return shipping.

Optical comparator mirror replacement

Warning: Disconnect all electrical power to the comparator before you begin.

Remove the optical comparator screen to allow access to the mirror. To do this, insert a hex wrench or other object into the center of the brass wheel at the top of the screen and lift the wheel up above the screen edge. While holding up on the wheel press in on the bottom of the optical comparator screen. This will allow the top to move out past the brass wheel. You can now gently release the brass wheel spring pressure and lift out the screen.

Lift out the three black hole covers in the rear of the comparator. This will allow you to access the optical comparator mirror mounting screws. Wrap tissue paper around the 12” mirror so it does not get damaged while removing it. Hold the mirror inside the optical comparator while removing the three screws in the rear of the machine  (two people are recommended for this procedure). The mirror is glued to a mirror frame. There are three springs between the mirror frame and the mirror supports inside the machine. Each of the mounting screws runs through a spring and into the mirror frame. When you remove the screws and the mirror, the springs will fall out. You should remove the screws and springs one at a time to prevent them from falling and damaging the lower mirror. You should protect the lower mirror from damage by wrapping the lower mirror with tissue paper and placing a sheet of cardboard in the machine between the two mirrors.

Installing the new optical comparator mirror

Hold a spring against the back of the mirror frame. Hold the mirror in place inside the machine while someone starts the screw from behind the comparator. It might be necessary to compress the spring slightly by hand to get the screw to start in the threads of the mirror frame. Start with the top screw then do the same for the bottom two. After starting all of the screws, tighten the screws evenly using a plastic gage (provided) or gage blocks (approximately .570”). Insert the gage between the mirror frame and the mirror mounting rails inside the optical comparator. Insert the gage near each of the springs. With the gage in place tighten each of the screws so that some resistance can be felt when the gage is removed. You want to get the mirror even on all three points. This will provide you with a good starting point when re-setting the magnification.

Install the screen by first setting it in the screen drive wheel (bottom right) while pressing back the small encoder wheel (lower right) with the edge of the screen. Set the bottom left of the screen in the left lower screen roller. Then while holding the screen close to the top brass roller lift the roller as you did for removal and push the screen into place. Let the roller down gently. Making sure the edge of the glass is seated in the roller groove. You can press in on the bottom of the screen to move the top of the screen out. This will help you when setting the roller down on the edge of the screen.

Caution: If the screen is not seated in the brass roller groove the screen might break when it is rotated and the roller snaps into place.

Note: Changing the optical comparator mirror will affect the magnification and the position of the X-axis stage horizon shown on the screen.

You should now be able to regain the correct magnification and horizon position by only adjusting the one mirror that you replaced. Set the horizontal line of the optical comparator screen to its proper position using the vernier scale and screen protractor. With the screen in the proper orientation bring the standard into correct magnification by adjusting the three screws in the rear of the machine. When resetting the magnification be sure to bring the horizon into its correct orientation on the screen. Adjusting the lower screws of the 12” mirror unevenly will affect the position of the X-axis horizon relative to the optical comparator screen.

Optical comparator internal edge adjustment

To adjust the optical comparator internal edge sensor so that it is exactly in the center of the screen cross lines use the following procedure.

Using only the optical comparator profile light, place an object on the optical comparator stage that will cast a straight image on the screen in a vertical and horizontal direction. A gage block would work fine. The image should cover all or at least half of the screen diameter.

Rotate the screen so the protractor is at zero. This will make the cross lines vertical and horizontal.

“Teach” the edge detector (refer to the optical comparator digital readout manual if necessary) and make sure that the readout beeps when you cross an edge. Note: You might have to move the image far from the center of the screen to test for the beep.

Bring the top of the image to the horizontal cross line. Making sure some of the image is left and right of the vertical cross line.

The adjustment for the optical comparator edge detector is located on the top of the comparator case. There are three button head screws in a triangle pattern approximately in the center on the case. Two screws on one side and one screw on the other. Turning either one of the screws on the side with two screws will move the location of the sensor up and down  (only a small amount of adjustment should be needed).

Adjust the up and down until the beep noise flutters. That indicates you are right on the line. Test the location by moving the image up and down. Re-adjust if necessary.

After the horizontal is adjusted align the right side of the image along the vertical cross line. Making sure some of the image is above and below the horizontal line. Adjust the left and right sensor using the single screw until the beep noise flutters. That indicates you are right on the line. Test the location by moving the image left and right. Re-adjust if necessary.

Danger : Disconnect all power to optical comparator and accessories before starting to work.

This repair should be done by a qualified service person. Please read through all instructions before beginning work.

Before removing the optical comparator fan note the direction of the airflow or if the fan is not working. Look on the fan for an arrow either stamped in the metal or a label indicating the direction of flow. During this replacement you will need to orient the new fan so the airflow is in the same direction. Disconnect the three wires going to the profile fan (there should be three round connectors about 6” from the fan, the wires can be separated at this point). Remove the screws holding the old profile fan in place and mount the new fan in place with the airflow in the correct direction. The new optical comparator fan has 5-millimeter threads for mounting or 4-millimeter screws can pass through the threads. Depending on the mounting style used for your optical comparator. Connect the two power wires and green ground wire of the new fan (it does not make a difference which of the black fan wires goes to which of the black and white supply wires). After the fan installation is complete reconnect the power to the optical comparator and accessories. The new fan should run when the machine power switch is turned on.

Optical comparator maintenance

Optical comparators must receive proper care to ensure longer lifespan and retain measurement accuracy. There are few parts that require regular cleaning:

  • lenses
  • mirrors
  • screens

Any accumulated dust must be removed from the surface of the lenses with a photographic lens brush. The surface of the inner optical comparator lens (projection or condenser) must not be touched. The element of the projection lens facing the mirror is most critical in this respect. A single finger mark on this surface will noticeably degrade the quality of the projected image. It should be noted that the projection lens element facing the work piece is far less critical.

Any grease or finger marks may be removed with a new photographic cleaning tissue. If necessary, the tissue may be moistened with methanol or lens cleaning fluid to assist cleaning.

Access to the optical comparator mirrors is gained by removing the screen. To remove grease or finger marks from the mirror surfaces use a surgical cotton wool swab soaked with methanol. Apply the swab very lightly to the mirror surface finally giving a light polish with another clean dry swab. Do not use force. All mirror surfaces should be cleaned as infrequently as possible. Never use polishes, window cleaning aerosols or liquids when cleaning the mirrors or lenses.

All optical comparator screens are made out of ground glass with printed cross lines. The cross lines can be removed if the screen is not cleaned properly. We recommend that the screens be cleaned only with soap (a mild non-lotion dish-washing solution) and plenty of water. Use a lint free soft cotton cloth to gently clean. Let the screen air dry.

If your environment is extremely dirty, please contact us about optional encapsulated image screens which cover the markings with a protective membrane.

Optical comparators operating basics part 2

Surface Illumination

Optical comparators illumination of surface features on solid objects is achieved through the use of high intensity fiber optic light guides. These guides are adjustable and can be extended by pulling on the cable ends protruding from either side of the lens mounts. They may be secured in place with the clamping screws on the side of each of the guide blocks.

For the brightest reflected image, place the light guides as close as possible to the face of the component to be inspected. Where components have a directional surface finish, e.g. ground faces, the surface texture should be placed at right angles to the table, this will give the brightest reflected image on the screen.

As the fiber optic cable is a “cold light” source, the heat radiated to the component under test is minimal.

Manual Table Adjustment

The X axis of the stage has a quick release for coarse motion. This feature should not be used to move the stage during a measuring routine. Also, when the quick release is engaged, the control knob of the X axis is disabled.

To engage the quick release, rotate the lever to the furthest position to the right. To re-engage the control knob rotate the lever to the furthest left position.

Stage Helix Adjustment

The top plate of the optical comparator stage has a helix adjustment, which allows the top plate of the stage to be rotated. To do this the 2 locking clamps located beneath the stage must first be loosened. The stage plate can now be rotated ±15 degrees. The helix indicator shows the amount of rotation in 5’ increments. Once the desired helix is adjusted tighten the locking clamps to hold this position.
To restore the helix to the zero (0) position, loosen the locking clamp rotate the helix adjustment to zero.

Vertical Stage Adjustment

Vertical adjustment of the optical comparator work stage is by means of the large hand wheel on the left below the work stage knee.

Screen Angular Measurement

The electronic venire protractor screen, consists of a frosted glass screen divided into four equal quadrants by precision cross lines. The screen is rotated by means of a small hand wheel on the right of the screen. The standard electronic venire protractor can then be used to determine angles with 1 minute resolution.

Electronic Rotary Protractor

16H optical comparators are equipped with a digital electronic rotary screen protractor. These compact units display the current rotary position of the screen and utilizes non-volatile memory to permanently store the calibration data regardless of power loss. Optical comparators display rotary position in either degrees and minutes or degrees and 1/100 of a degree.

The guide below applies to Dorsey Metrology 16H Benchmark optical comparator. It might not reflect actual features found on other products.

First step in optical comparator operation is of course switching the power on. The projector lamps are usually controlled by rocker switches on the lower front panel of the projector body. Separate switches control the profile and surface illumination lamps. The profile lamp uses a 2 position switch: down is low intensity, while up is high intensity.

Before using optical comparator to inspect part some pre-run checks might be required. First ensure that the projector is secure on its stand or on a substantial bench. Having selected a lens of the required magnification and locked it securely in position, locate the component to be inspected in a suitable work holding fixture which must be securely clamped to the table, the optical comparator is now ready for use.

Focusing

Focus is achieved by rotating the small hand wheel on the right hand side of the stage knee. The lens focal plane is approximately mid-way along the focus traverse, central to the dovetail slot in the table.

Profile Projection Intensity

There are two profile light intensity levels, high or low, selected from the rocker switch on the panel below the screen. Select the level that gives the most comfortable viewing, according to the object projected and the external lighting conditions prevailing. Profile lighting is provided by a 24V 150W tungsten halogen lamp. An iris diaphragm and or green filter is available as an optional extra which may be fitted direct onto the lamp house to reduce the level of reflection from brightly polished surfaces e.g. ground threads, etc.

Optical comparators safety

Optical comparators are designed and manufactured so far as is reasonably practical to allow its safe operation when used in accordance with the safety instructions. Optical comparator must be used in position that does not constitute a hazard, where the operator and maintenance staff have free access to the control and maintenance of the equipment, and are not subject to any external hazards. Services should be conducted by an authorized person, to approved safety standards. The electrical supply should be taken from a correctly rated source.

Power requirements

The standard optical comparator is usually supplied to accept 110-120V AC, 50-60 Hz main power. It may also be configured to accept either 220-240V AC, 50-60HZ or 100V AC, 50-60Hz. Unless clearly labeled (next to the power input plug) optical comparators are configured for 110-120V AC, 50-60Hz. Listed below are wire color codes:

Wire color codes
110-120V AC system Brown/Black Hot
Blue/White Neutral
Green Ground
220-240V AC system Brown/Black Hot 1
Blue/White Hot 2
Green Ground

Note optical comparators must be plugged into an outlet rated at a minimum of 15A and must not be operated without an adequate ground connection.