Welcome to Optical Comparators blog.

Since 1986 Dorsey Metrology have been manufacturing high quality optical comparators.

You can contact us about optical comparators sales, service and general informaion by calling directly to +1-845-454-3111.

Tag: manuals

Optical comparator internal edge adjustment

To adjust the optical comparator internal edge sensor so that it is exactly in the center of the screen cross lines use the following procedure.

Using only the optical comparator profile light, place an object on the optical comparator stage that will cast a straight image on the screen in a vertical and horizontal direction. A gage block would work fine. The image should cover all or at least half of the screen diameter.

Rotate the screen so the protractor is at zero. This will make the cross lines vertical and horizontal.

“Teach” the edge detector (refer to the optical comparator digital readout manual if necessary) and make sure that the readout beeps when you cross an edge. Note: You might have to move the image far from the center of the screen to test for the beep.

Bring the top of the image to the horizontal cross line. Making sure some of the image is left and right of the vertical cross line.

The adjustment for the optical comparator edge detector is located on the top of the comparator case. There are three button head screws in a triangle pattern approximately in the center on the case. Two screws on one side and one screw on the other. Turning either one of the screws on the side with two screws will move the location of the sensor up and down  (only a small amount of adjustment should be needed).

Adjust the up and down until the beep noise flutters. That indicates you are right on the line. Test the location by moving the image up and down. Re-adjust if necessary.

After the horizontal is adjusted align the right side of the image along the vertical cross line. Making sure some of the image is above and below the horizontal line. Adjust the left and right sensor using the single screw until the beep noise flutters. That indicates you are right on the line. Test the location by moving the image left and right. Re-adjust if necessary.

Danger : Disconnect all power to optical comparator and accessories before starting to work.

This repair should be done by a qualified service person. Please read through all instructions before beginning work.

Before removing the optical comparator fan note the direction of the airflow or if the fan is not working. Look on the fan for an arrow either stamped in the metal or a label indicating the direction of flow. During this replacement you will need to orient the new fan so the airflow is in the same direction. Disconnect the three wires going to the profile fan (there should be three round connectors about 6” from the fan, the wires can be separated at this point). Remove the screws holding the old profile fan in place and mount the new fan in place with the airflow in the correct direction. The new optical comparator fan has 5-millimeter threads for mounting or 4-millimeter screws can pass through the threads. Depending on the mounting style used for your optical comparator. Connect the two power wires and green ground wire of the new fan (it does not make a difference which of the black fan wires goes to which of the black and white supply wires). After the fan installation is complete reconnect the power to the optical comparator and accessories. The new fan should run when the machine power switch is turned on.

Optical comparator maintenance

Optical comparators must receive proper care to ensure longer lifespan and retain measurement accuracy. There are few parts that require regular cleaning:

  • lenses
  • mirrors
  • screens

Any accumulated dust must be removed from the surface of the lenses with a photographic lens brush. The surface of the inner optical comparator lens (projection or condenser) must not be touched. The element of the projection lens facing the mirror is most critical in this respect. A single finger mark on this surface will noticeably degrade the quality of the projected image. It should be noted that the projection lens element facing the work piece is far less critical.

Any grease or finger marks may be removed with a new photographic cleaning tissue. If necessary, the tissue may be moistened with methanol or lens cleaning fluid to assist cleaning.

Access to the optical comparator mirrors is gained by removing the screen. To remove grease or finger marks from the mirror surfaces use a surgical cotton wool swab soaked with methanol. Apply the swab very lightly to the mirror surface finally giving a light polish with another clean dry swab. Do not use force. All mirror surfaces should be cleaned as infrequently as possible. Never use polishes, window cleaning aerosols or liquids when cleaning the mirrors or lenses.

All optical comparator screens are made out of ground glass with printed cross lines. The cross lines can be removed if the screen is not cleaned properly. We recommend that the screens be cleaned only with soap (a mild non-lotion dish-washing solution) and plenty of water. Use a lint free soft cotton cloth to gently clean. Let the screen air dry.

If your environment is extremely dirty, please contact us about optional encapsulated image screens which cover the markings with a protective membrane.

Optical comparators operating basics part 2

Surface Illumination

Optical comparators illumination of surface features on solid objects is achieved through the use of high intensity fiber optic light guides. These guides are adjustable and can be extended by pulling on the cable ends protruding from either side of the lens mounts. They may be secured in place with the clamping screws on the side of each of the guide blocks.

For the brightest reflected image, place the light guides as close as possible to the face of the component to be inspected. Where components have a directional surface finish, e.g. ground faces, the surface texture should be placed at right angles to the table, this will give the brightest reflected image on the screen.

As the fiber optic cable is a “cold light” source, the heat radiated to the component under test is minimal.

Manual Table Adjustment

The X axis of the stage has a quick release for coarse motion. This feature should not be used to move the stage during a measuring routine. Also, when the quick release is engaged, the control knob of the X axis is disabled.

To engage the quick release, rotate the lever to the furthest position to the right. To re-engage the control knob rotate the lever to the furthest left position.

Stage Helix Adjustment

The top plate of the optical comparator stage has a helix adjustment, which allows the top plate of the stage to be rotated. To do this the 2 locking clamps located beneath the stage must first be loosened. The stage plate can now be rotated ±15 degrees. The helix indicator shows the amount of rotation in 5’ increments. Once the desired helix is adjusted tighten the locking clamps to hold this position.
To restore the helix to the zero (0) position, loosen the locking clamp rotate the helix adjustment to zero.

Vertical Stage Adjustment

Vertical adjustment of the optical comparator work stage is by means of the large hand wheel on the left below the work stage knee.

Screen Angular Measurement

The electronic venire protractor screen, consists of a frosted glass screen divided into four equal quadrants by precision cross lines. The screen is rotated by means of a small hand wheel on the right of the screen. The standard electronic venire protractor can then be used to determine angles with 1 minute resolution.

Electronic Rotary Protractor

16H optical comparators are equipped with a digital electronic rotary screen protractor. These compact units display the current rotary position of the screen and utilizes non-volatile memory to permanently store the calibration data regardless of power loss. Optical comparators display rotary position in either degrees and minutes or degrees and 1/100 of a degree.

The guide below applies to Dorsey Metrology 16H Benchmark optical comparator. It might not reflect actual features found on other products.

First step in optical comparator operation is of course switching the power on. The projector lamps are usually controlled by rocker switches on the lower front panel of the projector body. Separate switches control the profile and surface illumination lamps. The profile lamp uses a 2 position switch: down is low intensity, while up is high intensity.

Before using optical comparator to inspect part some pre-run checks might be required. First ensure that the projector is secure on its stand or on a substantial bench. Having selected a lens of the required magnification and locked it securely in position, locate the component to be inspected in a suitable work holding fixture which must be securely clamped to the table, the optical comparator is now ready for use.

Focusing

Focus is achieved by rotating the small hand wheel on the right hand side of the stage knee. The lens focal plane is approximately mid-way along the focus traverse, central to the dovetail slot in the table.

Profile Projection Intensity

There are two profile light intensity levels, high or low, selected from the rocker switch on the panel below the screen. Select the level that gives the most comfortable viewing, according to the object projected and the external lighting conditions prevailing. Profile lighting is provided by a 24V 150W tungsten halogen lamp. An iris diaphragm and or green filter is available as an optional extra which may be fitted direct onto the lamp house to reduce the level of reflection from brightly polished surfaces e.g. ground threads, etc.

Optical comparators safety

Optical comparators are designed and manufactured so far as is reasonably practical to allow its safe operation when used in accordance with the safety instructions. Optical comparator must be used in position that does not constitute a hazard, where the operator and maintenance staff have free access to the control and maintenance of the equipment, and are not subject to any external hazards. Services should be conducted by an authorized person, to approved safety standards. The electrical supply should be taken from a correctly rated source.

Power requirements

The standard optical comparator is usually supplied to accept 110-120V AC, 50-60 Hz main power. It may also be configured to accept either 220-240V AC, 50-60HZ or 100V AC, 50-60Hz. Unless clearly labeled (next to the power input plug) optical comparators are configured for 110-120V AC, 50-60Hz. Listed below are wire color codes:

Wire color codes
110-120V AC system Brown/Black Hot
Blue/White Neutral
Green Ground
220-240V AC system Brown/Black Hot 1
Blue/White Hot 2
Green Ground

Note optical comparators must be plugged into an outlet rated at a minimum of 15A and must not be operated without an adequate ground connection.